Near infrared spectrometer
Publish:Box Optronics  Time:2024-03-26  Views:95
The near-infrared spectrum is mainly generated when the molecular vibration transitions from the ground state to a high energy level due to the non-resonant nature of the molecular vibration. What is recorded is mainly the frequency doubling and combined frequency absorption of the vibration of the hydrogen-containing group X-H (X=C, N, O). . Different groups (such as methyl, methylene, benzene rings, etc.) or the same group have obvious differences in the near-infrared absorption wavelength and intensity in different chemical environments.
Die Nahinfrarot-Spektroskopie verfügt über reiche Struktur- und Kompositionsinformationen und eignet sich sehr gut zur Messung der Zusammensetzung und Eigenschaften von organischen Kohlenwasserstoffen. Im Nahinfrarotspektrumbereich ist die Absorptionsintensität jedoch schwach, die Empfindlichkeit ist relativ niedrig, und die Absorptionsbänder sind breit und überlappen sich ernsthaft. Daher ist es sehr schwierig, quantitative Analysen durchzuführen, die sich auf die traditionelle Methode zur Festlegung einer Arbeitskurve stützen. Die Entwicklung der Chemometrie hat eine mathematische Grundlage für die Lösung dieses Problems gelegt. Es funktioniert nach dem Prinzip, dass, wenn die Zusammensetzung der Probe gleich ist, ihr Spektrum das gleiche ist und umgekehrt. Wenn wir die Korrespondenz zwischen dem Spektrum und den zu messenden Parametern feststellen (sogenanntes analytisches Modell), dann können, solange das Spektrum der Probe gemessen wird, die erforderlichen Qualitätsparameter-Daten schnell durch das Spektrum und die obige Korrespondenz erhalten werden.
How to measure near infrared spectroscopy
Like conventional molecular absorption spectrometry analysis, measuring the transmission spectrum of solution samples in near-infrared spectroscopy technology is one of its main measurement methods. In addition, it is also commonly used to directly measure the diffuse reflectance spectrum of solid samples, such as flakes, granules, powders, and even viscous liquid or paste samples. In the field of near-infrared spectroscopy, commonly used measurement methods include transmission, diffuse reflection, diffuse transmission, and transflectance.
1. Transmission mode
Like other molecular absorption spectra, the measurement of near-infrared transmission spectrum is used for clear, transparent and uniform liquid samples. The most commonly used measurement accessory is a quartz cuvette, and the measurement index is absorbance. The relationship between spectral absorbance, optical path length and sample concentration is consistent with Lambert-Beer‘s law, that is, the absorbance is directly proportional to the optical path length and sample concentration. This is the basis for quantitative analysis of near-infrared spectroscopy.
The sensitivity of near-infrared spectroscopy is very low, so it is generally not necessary to dilute the sample during analysis. However, solvents, including water, have obvious absorption of near-infrared light. When the optical path of the cuvette is too large, the absorbance will be very high, even saturated. Therefore, in order to reduce analysis errors, the absorbance of the measured spectrum is best controlled between 0.1-1, and cuvettes of 1-10 mm are generally used. Sometimes for convenience, near-infrared spectroscopy measurements with absorbance as low as 0.01, or as high as 1.5, or even 2 are often seen.
2. Diffuse reflection mode
The outstanding advantages of near-infrared spectroscopy technology, such as non-destructive measurement, no need for sample preparation, simplicity and speed, etc., mainly stem from its diffuse reflection spectrum collection mode. The diffuse reflection mode can be used for the measurement of solid samples such as powders, blocks, sheets, and silk, as well as semi-solid samples such as pastes and pastes. The sample can be in any shape, such as fruit, tablets, cereals, paper, dairy, meat, etc. No special sample preparation is required and can be measured directly.
Near-infrared diffuse reflection spectrum does not comply with Lambert-Beer‘s law, but previous studies have found that the absorbance of diffuse reflection (actually the negative logarithm of the ratio of sample reflectance to reference reflectance) and concentration have a certain relationship under certain conditions. For a linear relationship, the conditions that need to be met include the sample thickness being large enough, the concentration range being narrow, the physical state of the sample and the spectral measurement conditions being consistent, etc. Therefore, using diffuse reflectance spectroscopy can also be used for quantitative analysis using multivariate correction like transmission spectroscopy.
4. Transflective mode
The transmission spectrum measurement of a solution sample is to pass the incident light through the sample and measure the transmission spectrum on the other side. Different from this, in the transflective mode, a reflective mirror is placed behind the sample solution. The incident light passes through the sample and is reflected by the mirror before entering the sample solution again.The transflective spectrum is measured on the same side of the incident light. Light passes through the sample twice, so the optical path length is twice that of a normal transmission spectrum. The transflective mode is designed for the convenience of measuring spectra. Because the incident light and reflected light are on the same side, you can install both the incident light path and the reflected light path in one probe, and install a cavity at the front end of the probe. The top is a reflector. When in use, the probe is inserted into the solution, the solution enters the cavity, the light shines into the solution from the incident light path, is reflected back to the solution on the reflector, and then enters the reflected light path and enters the spectrometer to measure the spectrum. In essence, the transmission and reflection spectrum is also a transmission spectrum, so its absorbance has a linear relationship with concentration.